Mouse DPD(Deoxypyridinoline) ELISA Kit

    • competitive-ELISA-Elabscience
    • competitive-ELISA-Elabscience
    • competitive-ELISA-Elabscience
    <
    >
    • competitive-ELISA-Elabscience
    • competitive-ELISA-Elabscience
    • competitive-ELISA-Elabscience
    • competitive-ELISA-Elabscience
    • competitive-ELISA-Elabscience
    • competitive-ELISA-Elabscience

      Catalog number:E-EL-M0408

      Synonyms:Pyrilinks-D, deoxyPYD

      Size:
      • 96T
      • 24T
      Qty:
      - +
      Price: $580

      Reactivity: Mouse

      Detection Range: 3.13~200 ng/mL

      Sensitivity: 1.88 ng/mL

      Lead Time: 7~10 daysWelcome to order from local distributors.

      Add to cart Compare Bulk request Manual

      Intended use

      This ELISA kit applies to the in vitro quantitative determination of Mouse DPD concentrations in serum, plasma and other biological fluids.

      Test principle

      This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Mouse DPD. During the reaction, Mouse DPD in the sample or standard competes with a fixed amount of Mouse DPD on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Mouse DPD. Excess conjugate and unbound sample or standard are washed from the plate, and Avidin conjugated to Horseradish Peroxidase (HRP) are added to each microplate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Mouse DPD in the samples is then determined by comparing the OD of the samples to the standard curve.

      Assay type Competitive-ELISA
      Format 96T
      Assay time 2.5h
      Reactivity Mouse
      Detection Method Colormetric
      Detection Range 3.13—200 ng/mL
      Sensitivity 1.88 ng/mL
      Sample Volume 50μL
      Sample Type Serum, plasma and other biological fluids

      Specificity

      This kit recognizes Mouse DPD in samples. No significant cross-reactivity or interference between Mouse DPD and analogues was observed.

      Typical data

      As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

      Concentration(ng/mL) O.D Average
      200 0.331
      0.341
      0.336
      100 0.436
      0.444
      0.44
      50 0.633
      0.621
      0.627
      25 0.899
      0.935
      0.917
      12.5 1.291
      1.275
      1.283
      6.25 1.648
      1.634
      1.641
      3.13 1.908
      1.918
      1.913
      0 2.268
      2.278
      2.273

      Precision

      Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Mouse DPD were tested 20 times on one plate, respectively.
      Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Mouse DPD were tested on 3 different plates, 20 replicates in each plate.

        Intra-assay Precision Inter-assay Precision
      Sample 1 2 3 1 2 3
      n 20 20 20 20 20 20
      Mean (ng/mL) 10.90 28.60 91.80 11.10 30.30 87.20
      Standard deviation 0.80 1.50 3.70 0.70 1.20 4.40
      C V (%) 7.34 5.24 4.03 6.31 3.96 5.05

      Recovery

      The recovery of Mouse DPD spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

      Sample Type Range (%) Average Recovery (%)
      Serum (n=5) 95-106 100
      EDTA plasma (n=5) 86-99 92
      Cell culture media (n=5) 84-98 91

      Linearity

      Samples were spiked with high concentrations of Mouse DPD and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

          Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
      1:2 Range (%) 94-105 86-101 100-115
      Average (%) 99 93 106
      1:4 Range (%) 87-97 89-103 95-110
      Average (%) 92 95 102
      1:8 Range (%) 88-102 93-107 98-109
      Average (%) 93 99 103
      1:16 Range (%) 87-102 93-108 94-106
      Average (%) 94 99 100

      Kit components & Storage

      An unopened kit can be stored at 4℃ for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.

      Item Specifications Storage
      Micro ELISA Plate(Dismountable) 8 wells ×12 strips -20℃, 6 months
      Reference Standard 2 vials
      Concentrated Biotinylated Detection Ab (100×) 1 vial, 120 μL
      Concentrated HRP Conjugate (100×) 1 vial, 120 μL -20℃(shading light), 6 months
      Reference Standard & Sample Diluent 1 vial, 20 mL 4℃, 6 months
      Biotinylated Detection Ab Diluent 1 vial, 14 mL
      HRP Conjugate Diluent 1 vial, 14 mL
      Concentrated Wash Buffer (25×) 1 vial, 30 mL
      Substrate Reagent 1 vial, 10 mL 4℃(shading light)
      Stop Solution 1 vial, 10 mL 4℃
      Plate Sealer 5 pieces
      Product Description 1 copy
      Certificate of Analysis 1 copy

      Note: All reagent bottle caps must be tightened to prevent evaporation and microbial pollution.
      The volume of reagents in partial shipments is a little more than the volume marked on the label, please use accurate measuring equipment instead of directly pouring into the vial(s).

      Other supplies required

      • Microplate reader with 450 nm wavelength filter
      • High-precision transfer pipette, EP tubes and disposable pipette tips
      • Incubator capable of maintaining 37℃
      • Deionized or distilled water
      • Absorbent paper
      • Loading slot for Wash Buffer

      Assay Procedure

      1.Add 50 μL standard or sample to each well.

      2.Immediately add 50 μL Biotinylated Detection Ab to each well.

      3.Incubate for 45 min at 37℃. Aspirate and wash 3 times.

      4.Add 100 μL HRP Conjugate to each well. Incubate for 30 min at 37℃. Aspirate and wash 5 times.

      5.Add 90 μL Substrate Reagent. Incubate for 15 min at 37℃.

      6.Add 50 μL Stop Solution.

      7.Read at 450 nm immediately. Calculation of results.