Human FGF21(Fibroblast Growth Factor 21) ELISA Kit (E-EL-H0074)

Human FGF21(Fibroblast Growth Factor 21) ELISA Kit - 1
  • Human FGF21(Fibroblast Growth Factor 21) ELISA Kit - 1
  • Human FGF21(Fibroblast Growth Factor 21) ELISA Kit - 2
  • Human FGF21(Fibroblast Growth Factor 21) ELISA Kit - 3
All Size Price Qty
96T $ 609.00
- +
48T $ 487.00
- +
24T $ 150.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 18.75 pg/mL
Detection Range 31.25-2000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human FGF21 in samples. No significant cross-reactivity or interference between Human FGF21 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Colorimetric method, ELISA, Sandwich
Assay Type Sandwich-ELISA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃
Expiration Date 12 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
2000 2.265 2.351 2.308 2.236
1000 1.552 1.556 1.554 1.482
500 0.870 0.882 0.876 0.804
250 0.534 0.532 0.533 0.461
125 0.293 0.295 0.294 0.222
62.5 0.177 0.175 0.176 0.104
31.25 0.122 0.134 0.128 0.056
0 0.070 0.074 0.072 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 103.560 174.300 694.890 105.650 159.470 642.230
Standard deviation 6.260 11.190 32.520 8.750 10.540 56.520
CV(%) 6.040 6.420 4.680 8.280 6.610 8.800
Recovery
The recovery of Human FGF21 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Cell culture media (n=8) 88-102 94
Serum (n=8) 84-95 90
EDTA plasma (n=8) 84-98 90
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human FGF21 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 97-113 91-105 95-110
Average 103 98 102
1:4 Range 84-96 83-96 93-104
Average 91 88 99
1:8 Range 83-94 81-91 93-103
Average 89 86 98
Stability
Each kit batch is subjected to accelerated stability testing and real-time stability monitoring. Sample performance is evaluated after storage at 37 °C for 10 days to assess the impact of elevated temperature on assay reliability and reagent integrity.
/ Variation range of 37°C mean
concentration / 2-8°C mean
Sample 1(n=16) 95.82-98.79
Sample 2(n=16) 104.66-110.39
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human FGF21. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human FGF21 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human FGF21, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human FGF21. You can calculate the concentration of Human FGF21 in the samples by comparing the OD of the samples to the standard curve.
Gene ID 26291
Uniport ID Q9NSA1
Research Area Cancer , Cardiovascular , Signal Transduction
Human FGF21(Fibroblast Growth Factor 21) ELISA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

Species

{{relatedElisaPage.numTotal}} Results

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定