Current Position:Home>>Kits>>Metabolism Assays>> Bradford Protein Colorimetric Assay Kit
Cat.No.:E-BC-K168-S
Welcome to order from local distributors.
Protocols for metabolism assays
Add to cart Bulk requestFor research use only. Order now, ship in 3 days
Synonyms | TP |
Sample type | Serum,plasma,animal tissue |
Sensitivity | 0.026 mg/mL |
Detection range | 0.026-1.2 mg/mL |
Detection method | Colorimetric method |
Assay type | Quantitative |
Assay time | 50 min |
Precision | Average inter-assay CV: 2.300%Average intra-assay CV: 1.800% |
Other instruments required | Micropipettor, Vortex mixer |
Other reagents required | Normal saline (0.9% NaCl), PBS (0.01 M, pH 7.4) |
Storage | Reagent 1: 2-8℃, Reagent 2: RT |
Valid period | 12 months |
It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.026-1.2 mg/mL).
The recommended dilution factor for different samples is as follows (for reference only):
Sample type |
Dilution factor |
Human serum |
40-60 |
10% Mouse liver tissue homogenization |
8-12 |
Rat serum |
40-60 |
Note: The diluent is normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4).
PMID: 33256026 |
DOI: 10.5650/jos.ess22033 |
|
PMID: 35965085 |
Sample: serum |
D****aSubmitted [ Nov 06 2023 ]
Asked: Can i use cell lysate in microplate reader for this kit?
Reply
adminSubmitted [ Nov 06 2023 ]
Answered: 1. E-BC-K168-S kit can detect serum, plasma and animal tissue samples. E-BC-K168-M is recommended if you want to test cell samples. 2. Our kit is based on the principle of chemical reaction, and the components in the lysate are very complex, which may have an impact on the kit reaction system, so it's not recommended to treat cell samples with lysate. It is recommended that cell samples be broken by mechanical homogenization or ultrasonic homogenization.
M*****aSubmitted [ Oct 10 2022 ]
Asked: Is Bradford the same as Coomassie Brilliant blue method?
Reply
adminSubmitted [ Oct 10 2022 ]
Answered: Bradford is another statement of Coomassie Brilliant Blue. It's the same thing.
M*****aSubmitted [ Sep 27 2022 ]
Asked: Is Bradford the same as Coomassie Brilliant blue method?
Reply
adminSubmitted [ Sep 27 2022 ]
Answered: Bradford is another statement of Coomassie Brilliant Blue. It's the same thing.
M*****aSubmitted [ Sep 14 2022 ]
Asked: How long is the shelf life of this kit? What is the standard for this kit?
Reply
adminSubmitted [ Sep 14 2022 ]
Answered: The shelf life of this kit is 6 months and the standard is BSA.