Human ABCA1(ATP-binding cassette transporter A1) CLIA Kit (E-CL-H0424)

Human ABCA1(ATP-binding cassette transporter A1) CLIA Kit - 1
  • Human ABCA1(ATP-binding cassette transporter A1) CLIA Kit - 1
  • Human ABCA1(ATP-binding cassette transporter A1) CLIA Kit - 2
  • Human ABCA1(ATP-binding cassette transporter A1) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 18.75 pg/mL
Detection Range 31.25-2000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human ABCA1 in samples.No significant cross-reactivity or interference between Human ABCA1 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
2000 55004.000 65630.000 60317.000 60283
1000 23807.000 24895.000 24351.000 24317
500 12019.000 10539.000 11279.000 11245
250 5648.000 5782.000 5715.000 5681
125 2973.000 2735.000 2854.000 2820
62.5 1390.000 1356.000 1373.000 1339
31.25 653.000 705.000 679.000 645
0 33.000 35.000 34.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 95.550 293.750 912.590 99.940 276.530 980.130
Standard deviation 11.760 31.490 100.750 11.550 32.110 79.390
CV(%) 12.310 10.720 11.040 11.560 11.610 8.100
Recovery
The recovery of Human ABCA1 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 100-116 106
EDTA plasma (n=8) 94-106 100
Cell culture media (n=8) 87-102 93
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human ABCA1 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 101-119 93-111 91-103
Average 109 101 97
1:4 Range 91-105 97-109 100-114
Average 98 103 106
1:8 Range 97-110 88-99 89-102
Average 104 94 95
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human ABCA1. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human ABCA1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human ABCA1, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human ABCA1. You can calculate the concentration of Human ABCA1 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 19
Uniport ID O95477
Research Area Cancer , Cardiovascular , Metabolism , Neuroscience , Signal Transduction
Human ABCA1(ATP-binding cassette transporter A1) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定