Human ACTN2 (Actinin Alpha 2) CLIA Kit (E-CL-H0214)

Human ACTN2 (Actinin Alpha 2) CLIA Kit - 1
  • Human ACTN2 (Actinin Alpha 2) CLIA Kit - 1
  • Human ACTN2 (Actinin Alpha 2) CLIA Kit - 2
  • Human ACTN2 (Actinin Alpha 2) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 46.88 pg/mL
Detection Range 78.13-5000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human ACTN2 in samples.No significant cross-reactivity or interference between Human ACTN2 and analogues was observed
Recovery 80%-120%
Sample Type Tissue homogenates, cell lysates and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
5000 60060.000 69404.000 64732.000 64699
2500 33593.000 40187.000 36890.000 36857
1250 20437.000 19001.000 19719.000 19686
625 9987.000 10503.000 10245.000 10212
312.5 5283.000 5019.000 5151.000 5118
156.25 2593.000 2429.000 2511.000 2478
78.13 1442.000 1490.000 1466.000 1433
0 32.000 34.000 33.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 255.150 495.460 1831.610 235.730 520.100 1693.410
Standard deviation 32.070 41.720 152.760 20.580 40.100 187.460
CV(%) 12.570 8.420 8.340 8.730 7.710 11.070
Recovery
The recovery of Human ACTN2 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 96-111 104
EDTA plasma (n=8) 89-102 95
Cell culture media (n=8) 84-96 90
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human ACTN2 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 88-101 90-102 94-110
Average 93 97 101
1:4 Range 94-105 96-108 98-114
Average 99 102 105
1:8 Range 95-109 94-111 92-104
Average 103 101 98
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human ACTN2. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human ACTN2 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human ACTN2, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human ACTN2. You can calculate the concentration of Human ACTN2 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 88
Uniport ID P35609
Research Area Cancer , Cardiovascular , Signal Transduction
Human ACTN2 (Actinin Alpha 2) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定