Human ALDH(Aldehyde Dehydrogenase, Mitochondrial) CLIA Kit (E-CL-H0249)

Human ALDH(Aldehyde Dehydrogenase, Mitochondrial) CLIA Kit - 1
  • Human ALDH(Aldehyde Dehydrogenase, Mitochondrial) CLIA Kit - 1
  • Human ALDH(Aldehyde Dehydrogenase, Mitochondrial) CLIA Kit - 2
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 46.88 pg/mL
Detection Range 78.13-5000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human ALDM in samples.No significant cross-reactivity or interference between Human ALDM and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
5000 36318.000 34330.000 35324.000 35230
2500 16432.000 17214.000 16823.000 16729
1250 7460.000 7910.000 7685.000 7591
625 3601.000 3613.000 3607.000 3513
312.5 1916.000 1874.000 1895.000 1801
156.25 809.000 837.000 823.000 729
78.13 347.000 371.000 359.000 265
0 103.000 85.000 94.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 218.590 564.650 2070.200 243.230 530.280 2180.020
Standard deviation 13.590 25.420 113.340 11.980 24.870 123.110
CV(%) 5.390 4.920 5.130 5.040 4.950 5.220
Recovery
The recovery of Human ALDM was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 85-100 92
EDTA plasma (n=8) 86-96 90
Cell culture media (n=8) 97-108 105
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human ALDM and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 98-109 98-111 96-107
Average 101 104 100
1:4 Range 95-106 96-112 99-110
Average 103 104 103
1:8 Range 87-97 92-105 95-107
Average 91 98 102
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human ALDM. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human ALDM and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human ALDM, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human ALDM. You can calculate the concentration of Human ALDM in the samples by comparing the RLU value of the samples to the standard curve.
Human ALDH(Aldehyde Dehydrogenase, Mitochondrial) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定