Human ALDOC (Aldolase C, Fructose Bisphosphate) CLIA Kit (E-CL-H0893)

Human ALDOC (Aldolase C, Fructose Bisphosphate) CLIA Kit - 1
  • Human ALDOC (Aldolase C, Fructose Bisphosphate) CLIA Kit - 1
  • Human ALDOC (Aldolase C, Fructose Bisphosphate) CLIA Kit - 2
  • Human ALDOC (Aldolase C, Fructose Bisphosphate) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 0.09 ng/mL
Detection Range 0.16-10 ng/mL
Sample Volume 100 μL
Total Assay Time 4 h 30 min
Reactivity Human
Specificity This kit recognizes Human ALDOC in samples.No significant cross-reactivity or interference between Human ALDOC and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
ng/mL OD1 OD2 Mean OD Corrected OD
10 33002.000 33794.000 33398.000 33370
5 15942.000 16714.000 16328.000 16300
2.5 8474.000 7200.000 7837.000 7809
1.25 3747.000 3831.000 3789.000 3761
0.63 1917.000 1675.000 1796.000 1768
0.31 943.000 849.000 896.000 868
0.16 359.000 365.000 362.000 334
0 27.000 29.000 28.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(ng/mL) 0.490 1.190 4.960 0.490 1.180 4.790
Standard deviation 0.040 0.110 0.420 0.060 0.130 0.400
CV(%) 8.160 9.240 8.470 12.240 11.020 8.350
Recovery
The recovery of Human ALDOC was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 87-103 94
EDTA plasma (n=8) 95-111 103
Cell culture media (n=8) 92-106 100
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human ALDOC and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 98-110 98-112 99-111
Average 104 105 105
1:4 Range 90-105 89-99 102-117
Average 97 94 109
1:8 Range 97-114 100-115 96-107
Average 105 107 102
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human ALDOC. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human ALDOC and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human ALDOC, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human ALDOC. You can calculate the concentration of Human ALDOC in the samples by comparing the RLU value of the samples to the standard curve.
Human ALDOC (Aldolase C, Fructose Bisphosphate) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定