Human CEA (Carcinoembryonic Antigen) CLIA Kit (E-CL-H0492)

Human CEA (Carcinoembryonic Antigen)  CLIA Kit - 1
  • Human CEA (Carcinoembryonic Antigen)  CLIA Kit - 1
  • Human CEA (Carcinoembryonic Antigen)  CLIA Kit - 2
  • Human CEA (Carcinoembryonic Antigen)  CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 9.38 pg/mL
Detection Range 15.63-1000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human CEA/CD66 in samples.No significant cross-reactivity or interference between Human CEA/CD66 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
1000 52871.000 63677.000 58274.000 58246
500 23772.000 23860.000 23816.000 23788
250 11135.000 9649.000 10392.000 10364
125 4681.000 5603.000 5142.000 5114
62.5 2473.000 2437.000 2455.000 2427
31.25 1391.000 1207.000 1299.000 1271
15.63 698.000 742.000 720.000 692
0 27.000 29.000 28.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 46.370 87.730 407.800 45.360 90.580 371.720
Standard deviation 4.110 6.270 29.360 4.890 6.990 24.530
CV(%) 8.860 7.150 7.200 10.780 7.720 6.600
Recovery
The recovery of Human CEA was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 98-113 105
EDTA plasma (n=8) 94-105 99
Cell culture media (n=8) 95-108 103
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human CEA and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 98-112 94-107 95-109
Average 106 100 101
1:4 Range 92-105 93-106 97-111
Average 98 99 105
1:8 Range 84-96 97-109 95-108
Average 91 104 102
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human CEA/CD66. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human CEA/CD66 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human CEA/CD66, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human CEA/CD66. You can calculate the concentration of Human CEA/CD66 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 1048
Uniport ID P06731
Research Area Cancer , Immunology , Stem Cells , Tags And Cell Markers
Human CEA (Carcinoembryonic Antigen)  CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定