Human cTn-I/TNNI3 (cardiac Troponin I)CLIA Kit (E-CL-H0495)

Human cTn-I/TNNI3 (cardiac Troponin I)CLIA Kit - 1
  • Human cTn-I/TNNI3 (cardiac Troponin I)CLIA Kit - 1
  • Human cTn-I/TNNI3 (cardiac Troponin I)CLIA Kit - 2
  • Human cTn-I/TNNI3 (cardiac Troponin I)CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 46.88 pg/mL
Detection Range 78.13-5000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human cTn-I/TNNI3 in samples.No significant cross-reactivity or interference between Human cTn-I/TNNI3 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
5000 58295.000 71145.000 64720.000 64678
2500 34172.000 40586.000 37379.000 37337
1250 20353.000 18581.000 19467.000 19425
625 10241.000 10325.000 10283.000 10241
312.5 5490.000 4894.000 5192.000 5150
156.25 2776.000 2434.000 2605.000 2563
78.13 1338.000 1406.000 1372.000 1330
0 40.000 44.000 42.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 254.530 513.410 1903.300 257.460 557.970 1898.330
Standard deviation 24.890 60.020 156.640 30.050 44.300 184.520
CV(%) 9.780 11.690 8.230 11.670 7.940 9.720
Recovery
The recovery of Human cTn-I was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 100-116 107
EDTA plasma (n=8) 92-106 99
Cell culture media (n=8) 93-107 98
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human cTn-I and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 94-108 101-115 92-104
Average 100 106 97
1:4 Range 97-111 86-98 95-110
Average 102 92 101
1:8 Range 86-101 97-113 85-100
Average 92 105 91
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human cTn-I/TNNI3. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human cTn-I/TNNI3 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human cTn-I/TNNI3, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human cTn-I/TNNI3. You can calculate the concentration of Human cTn-I/TNNI3 in the samples by comparing the RLU value of the samples to the standard curve.
Human cTn-I/TNNI3 (cardiac Troponin I)CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定