Facebook
Toll-free:1-888-852-8623

All categories

  • All categories
  • Flow Cytometry Antibodies
  • Immunoassays
  • Antibodies and Reagents
  • Cell Health and Metabolism Assay Kits
  • Cell Isolation Products
  • Cell Identification Kits
  • Proteins and Peptides
  • Cell Culture
Please enter the item number/product keyword!
Keyword cannot be empty !
INSERT SYMBOLS:
  • α
  • β
  • γ
  • δ
  • ε
  • ζ
  • η
  • θ
  • κ
  • μ
  • ω
  • σ
  • τ
  • λ
  • ⅩⅢ
  • ⅩⅢ
  • ⅩⅣ
  • ⅩⅤ
  • ⅩⅦ
  • ⅩⅧ
  • UP ↑

Human CTSD (Cathepsin D) CLIA Kit (E-CL-H0506)

Human CTSD (Cathepsin D) CLIA Kit - 1
  • Human CTSD (Cathepsin D) CLIA Kit - 1
  • Human CTSD (Cathepsin D) CLIA Kit - 2
  • Human CTSD (Cathepsin D) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 9.38 pg/mL
Detection Range 15.63-1000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human CTSD in samples.No significant cross-reactivity or interference between Human CTSD and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
1000 56520.000 58908.000 57714.000 57688
500 21967.000 23417.000 22692.000 22666
250 11681.000 10351.000 11016.000 10990
125 4650.000 5164.000 4907.000 4881
62.5 2661.000 2309.000 2485.000 2459
31.25 1440.000 1304.000 1372.000 1346
15.63 702.000 758.000 730.000 704
0 25.000 27.000 26.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 47.800 135.430 338.170 49.780 136.960 323.890
Standard deviation 4.880 12.920 36.220 6.050 13.370 31.810
CV(%) 10.210 9.540 10.710 12.150 9.760 9.820
Recovery
The recovery of Human CTSD was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 99-117 107
EDTA plasma (n=8) 100-113 107
Cell culture media (n=8) 97-109 103
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human CTSD and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 90-104 99-113 94-105
Average 98 107 100
1:4 Range 89-99 87-101 97-113
Average 94 93 104
1:8 Range 92-105 91-106 90-106
Average 98 99 97
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human CTSD. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human CTSD and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human CTSD, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human CTSD. You can calculate the concentration of Human CTSD in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 1509
Uniport ID P07339
Research Area Cancer , Cell Biology , Neuroscience , Signal Transduction
Human CTSD (Cathepsin D) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定