Human GFAP (Glial Fibrillary Acidic Protein) CLIA Kit (E-CL-H1167)

Human GFAP (Glial Fibrillary Acidic Protein) CLIA Kit - 1
  • Human GFAP (Glial Fibrillary Acidic Protein) CLIA Kit - 1
  • Human GFAP (Glial Fibrillary Acidic Protein) CLIA Kit - 2
  • Human GFAP (Glial Fibrillary Acidic Protein) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 37.5 pg/mL
Detection Range 62.5-4000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human GFAP in samples.No significant cross-reactivity or interference between Human GFAP and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
4000 54336.000 65020.000 59678.000 59648
2000 27125.000 32775.000 29950.000 29920
1000 13566.000 13450.000 13508.000 13478
500 6306.000 7418.000 6862.000 6832
250 3612.000 3336.000 3474.000 3444
125 1768.000 1680.000 1724.000 1694
62.5 770.000 868.000 819.000 789
0 29.000 31.000 30.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 216.210 554.750 1858.100 196.810 548.780 1766.950
Standard deviation 23.240 55.250 139.730 19.620 39.900 158.670
CV(%) 10.750 9.960 7.520 9.970 7.270 8.980
Recovery
The recovery of Human GFAP was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 92-106 98
EDTA plasma (n=8) 94-106 101
Cell culture media (n=8) 88-100 93
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human GFAP and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 85-96 89-103 103-120
Average 91 96 110
1:4 Range 98-110 101-116 97-112
Average 104 107 104
1:8 Range 98-111 89-102 92-106
Average 105 96 100
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human GFAP. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human GFAP and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human GFAP, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human GFAP. You can calculate the concentration of Human GFAP in the samples by comparing the RLU value of the samples to the standard curve.
Human GFAP (Glial Fibrillary Acidic Protein) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定