Human IgG2 (Immunoglobulin G2) CLIA Kit (E-CL-H1093)

Human IgG2 (Immunoglobulin G2) CLIA Kit - 1
  • Human IgG2 (Immunoglobulin G2) CLIA Kit - 1
  • Human IgG2 (Immunoglobulin G2) CLIA Kit - 2
  • Human IgG2 (Immunoglobulin G2) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 56.25 ng/mL
Detection Range 93.75-6000 ng/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human IgG2 in samples.No significant cross-reactivity or interference between Human IgG2 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
ng/mL OD1 OD2 Mean OD Corrected OD
6000 66287.000 67203.000 66745.000 66713
3000 33403.000 37433.000 35418.000 35386
1500 20020.000 18758.000 19389.000 19357
750 8335.000 9671.000 9003.000 8971
375 4369.000 4347.000 4358.000 4326
187.5 1698.000 1588.000 1643.000 1611
93.75 450.000 496.000 473.000 441
0 31.000 33.000 32.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(ng/mL) 273.500 923.360 2626.870 267.850 939.040 2622.310
Standard deviation 28.530 82.640 196.750 32.360 71.180 181.730
CV(%) 10.430 8.950 7.490 12.080 7.580 6.930
Recovery
The recovery of Human IgG2 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 96-111 102
EDTA plasma (n=8) 98-110 103
Cell culture media (n=8) 88-102 95
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human IgG2 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 97-109 85-97 94-109
Average 103 92 101
1:4 Range 93-108 100-116 88-101
Average 101 107 94
1:8 Range 95-110 88-103 93-103
Average 100 94 98
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human IgG2. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human IgG2 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IgG2, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human IgG2. You can calculate the concentration of Human IgG2 in the samples by comparing the RLU value of the samples to the standard curve.
Uniport ID P01859
Research Area Immunology
Human IgG2 (Immunoglobulin G2) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定