Human IgG3 (Immunoglobulin G3) CLIA Kit (E-CL-H1094)

Human IgG3 (Immunoglobulin G3) CLIA Kit - 1
  • Human IgG3 (Immunoglobulin G3) CLIA Kit - 1
  • Human IgG3 (Immunoglobulin G3) CLIA Kit - 2
  • Human IgG3 (Immunoglobulin G3) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 18.75 pg/mL
Detection Range 31.25-2000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human IgG3 in samples.No significant cross-reactivity or interference between Human IgG3 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
2000 55895.000 66327.000 61111.000 61082
1000 23863.000 24141.000 24002.000 23973
500 12178.000 11816.000 11997.000 11968
250 5485.000 5729.000 5607.000 5578
125 2982.000 2634.000 2808.000 2779
62.5 1498.000 1286.000 1392.000 1363
31.25 670.000 762.000 716.000 687
0 28.000 30.000 29.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 94.630 243.990 961.690 97.630 229.600 1033.910
Standard deviation 11.550 20.350 105.310 12.510 24.200 90.360
CV(%) 12.210 8.340 10.950 12.810 10.540 8.740
Recovery
The recovery of Human IgG3 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 98-111 104
EDTA plasma (n=8) 93-105 98
Cell culture media (n=8) 100-116 107
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human IgG3 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 95-109 95-108 88-100
Average 102 100 94
1:4 Range 92-105 94-109 99-117
Average 100 101 107
1:8 Range 95-110 102-118 101-117
Average 101 108 107
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human IgG3. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human IgG3 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IgG3, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human IgG3. You can calculate the concentration of Human IgG3 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 7083
Uniport ID P01860
Research Area Immunology
Human IgG3 (Immunoglobulin G3) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定