Human MMP-3 (Matrix Metalloproteinase 3) CLIA Kit (E-CL-H0931)

Human MMP-3 (Matrix Metalloproteinase 3) CLIA Kit - 1
  • Human MMP-3 (Matrix Metalloproteinase 3) CLIA Kit - 1
  • Human MMP-3 (Matrix Metalloproteinase 3) CLIA Kit - 2
  • Human MMP-3 (Matrix Metalloproteinase 3) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 18.75 pg/mL
Detection Range 31.25-2000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human MMP-3 in samples.No significant cross-reactivity or interference between Human MMP-3 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
2000 56010.000 62748.000 59379.000 59340
1000 23320.000 25226.000 24273.000 24234
500 12430.000 10658.000 11544.000 11505
250 5536.000 6072.000 5804.000 5765
125 2808.000 2458.000 2633.000 2594
62.5 1396.000 1392.000 1394.000 1355
31.25 691.000 727.000 709.000 670
0 37.000 41.000 39.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 106.700 311.190 682.010 111.190 301.410 673.140
Standard deviation 9.020 30.530 48.970 10.160 22.360 74.920
CV(%) 8.450 9.810 7.180 9.140 7.420 11.130
Recovery
The recovery of Human MMP-3 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 101-113 107
EDTA plasma (n=8) 87-98 93
Cell culture media (n=8) 87-100 92
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human MMP-3 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 92-104 98-112 95-110
Average 98 105 103
1:4 Range 96-112 88-99 95-110
Average 102 94 101
1:8 Range 103-116 94-106 103-120
Average 110 99 109
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human MMP-3. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human MMP-3 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human MMP-3, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human MMP-3. You can calculate the concentration of Human MMP-3 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 4314
Uniport ID P08254
Research Area Cancer , Cell Biology , Cardiovascular , Signal Transduction
Human MMP-3 (Matrix Metalloproteinase 3) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定