Human NSE (Neuron-Specific Enolase) CLIA Kit (E-CL-H0702)

Human NSE (Neuron-Specific Enolase) CLIA Kit - 1
  • Human NSE (Neuron-Specific Enolase) CLIA Kit - 1
  • Human NSE (Neuron-Specific Enolase) CLIA Kit - 2
  • Human NSE (Neuron-Specific Enolase) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 0.28 ng/mL
Detection Range 0.47-30 ng/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human NSE in samples.No significant cross-reactivity or interference between Human NSE and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
ng/mL OD1 OD2 Mean OD Corrected OD
30 39007.000 39021.000 39014.000 38984
15 13182.000 15632.000 14407.000 14377
7.5 6343.000 5963.000 6153.000 6123
3.75 2859.000 3087.000 2973.000 2943
1.88 1665.000 1621.000 1643.000 1613
0.94 1086.000 930.000 1008.000 978
0.47 671.000 703.000 687.000 657
0 29.000 31.000 30.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(ng/mL) 1.490 4.370 10.890 1.350 4.660 11.500
Standard deviation 0.190 0.400 1.210 0.170 0.490 0.700
CV(%) 12.750 9.150 11.110 12.590 10.520 6.090
Recovery
The recovery of Human NSE was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 99-117 107
EDTA plasma (n=8) 95-110 100
Cell culture media (n=8) 89-102 95
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human NSE and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 93-107 99-112 93-103
Average 101 106 98
1:4 Range 96-110 96-107 86-101
Average 103 102 92
1:8 Range 94-111 101-117 91-106
Average 101 109 98
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human NSE. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human NSE and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human NSE, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human NSE. You can calculate the concentration of Human NSE in the samples by comparing the RLU value of the samples to the standard curve.
Human NSE (Neuron-Specific Enolase) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定