Human P-Cadherin (Cadherin, Placental) CLIA Kit (E-CL-H0624)

Human P-Cadherin (Cadherin, Placental) CLIA Kit - 1
  • Human P-Cadherin (Cadherin, Placental) CLIA Kit - 1
  • Human P-Cadherin (Cadherin, Placental) CLIA Kit - 2
  • Human P-Cadherin (Cadherin, Placental) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 9.38 pg/mL
Detection Range 15.63-1000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human P-Cadherin in samples.No significant cross-reactivity or interference between Human P-Cadherin and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
1000 50949.000 61731.000 56340.000 56308
500 20978.000 24260.000 22619.000 22587
250 11741.000 10035.000 10888.000 10856
125 4762.000 5364.000 5063.000 5031
62.5 2845.000 2489.000 2667.000 2635
31.25 1380.000 1196.000 1288.000 1256
15.63 695.000 759.000 727.000 695
0 31.000 33.000 32.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 52.300 132.830 408.680 56.360 141.500 442.940
Standard deviation 4.690 10.800 25.090 5.280 9.920 27.640
CV(%) 8.970 8.130 6.140 9.370 7.010 6.240
Recovery
The recovery of Human P-Cadherin was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 99-116 107
EDTA plasma (n=8) 91-106 98
Cell culture media (n=8) 98-112 105
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human P-Cadherin and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 86-98 98-114 89-103
Average 92 105 95
1:4 Range 88-100 99-114 92-107
Average 93 106 99
1:8 Range 97-109 93-108 86-102
Average 103 101 93
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human P-Cadherin. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human P-Cadherin and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human P-Cadherin, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human P-Cadherin. You can calculate the concentration of Human P-Cadherin in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 1001
Uniport ID P22223
Research Area Cancer , Neuroscience , Signal Transduction
Human P-Cadherin (Cadherin, Placental) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定