Human sVCAM-1/CD106(soluble Vascular Cell Adhesion Molecule 1) CLIA Kit (E-CL-H1287)

All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Reactivity Human
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
OD1 OD2 Mean OD Corrected OD
40 50262.000 49848.000 50055.000 49967
20 27298.000 27062.000 27180.000 27092
10 11021.000 10525.000 10773.000 10685
5 5319.000 5177.000 5248.000 5160
2.5 2837.000 2699.000 2768.000 2680
1.25 1440.000 1426.000 1433.000 1345
0.63 813.000 817.000 815.000 727
0 92.000 84.000 88.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean() 1.860 5.380 19.250 1.920 5.780 18.680
Standard deviation 0.120 0.250 0.830 0.100 0.290 0.980
CV(%) 6.610 4.640 4.320 5.070 4.940 5.250
Recovery
The recovery of Human sVCAM-1/CD106 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 90-103 97
EDTA plasma (n=8) 93-109 99
Cell culture media (n=8) 93-106 100
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human sVCAM-1/CD106 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 87-100 93-109 95-108
Average 93 99 100
1:4 Range 83-95 90-102 92-106
Average 89 96 99
1:8 Range 92-104 91-108 94-106
Average 99 98 100
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定