Human TNF-α (Tumor Necrosis Factor Alpha) CLIA Kit (E-CL-H0108)

Human TNF-α (Tumor Necrosis Factor Alpha) CLIA Kit - 1
  • Human TNF-α (Tumor Necrosis Factor Alpha) CLIA Kit - 1
  • Human TNF-α (Tumor Necrosis Factor Alpha) CLIA Kit - 2
  • Human TNF-α (Tumor Necrosis Factor Alpha) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 4.69 pg/mL
Detection Range 7.81-500 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human TNF-α in samples.No significant cross-reactivity or interference between Human TNF-α and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
500 51674.000 60026.000 55850.000 55813
250 19548.000 22946.000 21247.000 21210
125 10252.000 9832.000 10042.000 10005
62.5 4416.000 5264.000 4840.000 4803
31.25 2523.000 2195.000 2359.000 2322
15.63 1293.000 1249.000 1271.000 1234
7.81 735.000 819.000 777.000 740
0 37.000 37.000 37.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 25.600 58.030 169.010 27.580 54.450 158.500
Standard deviation 2.690 5.500 13.540 2.460 6.400 14.990
CV(%) 10.510 9.480 8.010 8.920 11.750 9.460
Recovery
The recovery of Human TNF-α was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 100-113 107
EDTA plasma (n=8) 92-103 98
Cell culture media (n=8) 87-101 92
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human TNF-α and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 97-110 98-112 86-99
Average 103 105 93
1:4 Range 85-98 101-117 99-111
Average 90 109 105
1:8 Range 93-106 91-105 90-102
Average 98 99 97
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human TNF-α. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human TNF-α and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human TNF-α, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human TNF-α. You can calculate the concentration of Human TNF-α in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 7124
Uniport ID P01375
Research Area Cancer , Cardiovascular , Metabolism , Immunology , Signal Transduction
Human TNF-α (Tumor Necrosis Factor Alpha) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定