Mouse IL-23 (Interleukin 23) CLIA Kit (E-CL-M0464)

Mouse IL-23 (Interleukin 23) CLIA Kit - 1
  • Mouse IL-23 (Interleukin 23) CLIA Kit - 1
  • Mouse IL-23 (Interleukin 23) CLIA Kit - 2
  • Mouse IL-23 (Interleukin 23) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 4.69 pg/mL
Detection Range 7.81-500 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Mouse
Specificity This kit recognizes Mouse IL-23 in samples.No significant cross-reactivity or interference between Mouse IL-23 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
500 50695.000 53903.000 52299.000 52271
250 20176.000 24432.000 22304.000 22276
125 10261.000 10025.000 10143.000 10115
62.5 4454.000 4650.000 4552.000 4524
31.25 2554.000 2466.000 2510.000 2482
15.63 1465.000 1255.000 1360.000 1332
7.81 731.000 779.000 755.000 727
0 27.000 29.000 28.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 23.640 58.910 176.630 24.050 53.320 177.190
Standard deviation 1.910 5.220 11.040 2.610 5.630 18.160
CV(%) 8.080 8.860 6.250 10.850 10.560 10.250
Recovery
The recovery of Mouse IL-23 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 97-111 104
EDTA plasma (n=8) 92-108 100
Cell culture media (n=8) 100-117 107
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Mouse IL-23 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 86-100 89-102 102-120
Average 93 96 109
1:4 Range 91-103 95-107 101-114
Average 97 102 107
1:8 Range 86-101 98-112 95-113
Average 93 104 103
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Mouse IL-23. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse IL-23 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse IL-23, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Mouse IL-23. You can calculate the concentration of Mouse IL-23 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 83430
Uniport ID Q9EQ14
Research Area Immunology
Mouse IL-23 (Interleukin 23) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定