Mouse Pro-ANP (Pro Atrial Natriuretic Peptide) CLIA Kit (E-CL-M0567)

Mouse Pro-ANP (Pro Atrial Natriuretic Peptide) CLIA Kit - 1
  • Mouse Pro-ANP (Pro Atrial Natriuretic Peptide) CLIA Kit - 1
  • Mouse Pro-ANP (Pro Atrial Natriuretic Peptide) CLIA Kit - 2
  • Mouse Pro-ANP (Pro Atrial Natriuretic Peptide) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 4.69 pg/mL
Detection Range 7.81-500 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Mouse
Specificity This kit recognizes Mouse Pro-ANP in samples.No significant cross-reactivity or interference between Mouse Pro-ANP and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
500 54897.000 57699.000 56298.000 56263
250 20690.000 21756.000 21223.000 21188
125 10105.000 9747.000 9926.000 9891
62.5 4447.000 4643.000 4545.000 4510
31.25 2564.000 2402.000 2483.000 2448
15.63 1388.000 1272.000 1330.000 1295
7.81 736.000 754.000 745.000 710
0 34.000 36.000 35.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 27.380 69.710 174.740 28.510 69.350 171.110
Standard deviation 2.980 6.950 13.930 3.230 8.220 10.850
CV(%) 10.880 9.970 7.970 11.330 11.850 6.340
Recovery
The recovery of Mouse Pro-ANP was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 89-105 96
EDTA plasma (n=8) 89-100 95
Cell culture media (n=8) 93-108 98
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Mouse Pro-ANP and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 84-96 96-107 99-115
Average 91 101 107
1:4 Range 98-114 94-108 89-100
Average 105 99 94
1:8 Range 102-114 85-98 89-103
Average 108 92 97
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Mouse Pro-ANP. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse Pro-ANP and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse Pro-ANP, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Mouse Pro-ANP. You can calculate the concentration of Mouse Pro-ANP in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 230899
Uniport ID P05125
Research Area Cancer , Cardiovascular , Signaltransduction
Mouse Pro-ANP (Pro Atrial Natriuretic Peptide) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定