Rat CRT (Calreticulin) CLIA Kit - 1
  • Rat CRT (Calreticulin) CLIA Kit - 1
  • Rat CRT (Calreticulin) CLIA Kit - 2
  • Rat CRT (Calreticulin) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 9.38 pg/mL
Detection Range 15.63-1000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Rat
Specificity This kit recognizes Rat CRT in samples.No significant cross-reactivity or interference between Rat CRT and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
1000 51524.000 59850.000 55687.000 55658
500 21513.000 26095.000 23804.000 23775
250 10476.000 10404.000 10440.000 10411
125 4787.000 5317.000 5052.000 5023
62.5 2581.000 2383.000 2482.000 2453
31.25 1409.000 1313.000 1361.000 1332
15.63 752.000 766.000 759.000 730
0 28.000 30.000 29.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 49.660 111.140 338.030 49.520 106.520 345.160
Standard deviation 4.480 10.170 32.890 5.460 11.650 35.760
CV(%) 9.020 9.150 9.730 11.030 10.940 10.360
Recovery
The recovery of Rat CRT was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 90-101 96
EDTA plasma (n=8) 89-103 95
Cell culture media (n=8) 91-104 97
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Rat CRT and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 90-102 93-105 91-104
Average 97 99 97
1:4 Range 103-118 98-115 96-110
Average 110 106 103
1:8 Range 101-117 89-100 85-95
Average 108 94 90
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Rat CRT. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Rat CRT and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Rat CRT, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Rat CRT. You can calculate the concentration of Rat CRT in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 64202
Uniport ID P18418
Research Area Cancer , Signaltransduction , TagsandCellMarkers
Rat CRT (Calreticulin) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定