Rat SCF (Stem Cell Factor) CLIA Kit (E-CL-R0618)

Rat SCF (Stem Cell Factor) CLIA Kit - 1
  • Rat SCF (Stem Cell Factor) CLIA Kit - 1
  • Rat SCF (Stem Cell Factor) CLIA Kit - 2
  • Rat SCF (Stem Cell Factor) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 4.69 pg/mL
Detection Range 7.81-500 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Rat
Specificity This kit recognizes Rat SCF in samples.No significant cross-reactivity or interference between Rat SCF and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
500 53697.000 56495.000 55096.000 55064
250 21690.000 23476.000 22583.000 22551
125 10216.000 9894.000 10055.000 10023
62.5 4387.000 4821.000 4604.000 4572
31.25 2432.000 2254.000 2343.000 2311
15.63 1376.000 1374.000 1375.000 1343
7.81 750.000 752.000 751.000 719
0 31.000 33.000 32.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 26.610 69.240 167.830 24.720 64.800 161.140
Standard deviation 3.430 5.030 16.250 2.700 4.700 18.310
CV(%) 12.890 7.260 9.680 10.920 7.250 11.360
Recovery
The recovery of Rat SCF was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 98-112 104
EDTA plasma (n=8) 94-106 100
Cell culture media (n=8) 99-113 106
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Rat SCF and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 94-107 98-115 98-114
Average 102 105 106
1:4 Range 95-109 99-117 99-115
Average 103 107 105
1:8 Range 101-111 84-98 96-112
Average 106 91 102
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Rat SCF. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Rat SCF and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Rat SCF, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Rat SCF. You can calculate the concentration of Rat SCF in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 60427
Uniport ID P21581
Research Area Cancer , Developmental Biology , Immunology , Stem Cells
Rat SCF (Stem Cell Factor) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定