Rat SDF-1α(Stromal Cell Derived Factor 1α)CLIA Kit (E-CL-R0751)

Rat SDF-1α(Stromal Cell Derived Factor 1α)CLIA Kit - 1
  • Rat SDF-1α(Stromal Cell Derived Factor 1α)CLIA Kit - 1
  • Rat SDF-1α(Stromal Cell Derived Factor 1α)CLIA Kit - 2
  • Rat SDF-1α(Stromal Cell Derived Factor 1α)CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 4.69 pg/mL
Detection Range 7.81-500 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Rat
Specificity This kit recognizes Rat SDF-1α in samples.No significant cross-reactivity or interference between Rat SDF-1α and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
500 51569.000 60645.000 56107.000 56073
250 20873.000 21919.000 21396.000 21362
125 10831.000 9667.000 10249.000 10215
62.5 4383.000 4741.000 4562.000 4528
31.25 2419.000 2215.000 2317.000 2283
15.63 1326.000 1238.000 1282.000 1248
7.81 730.000 790.000 760.000 726
0 33.000 35.000 34.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 23.440 57.840 230.700 21.510 54.780 213.080
Standard deviation 2.300 4.200 23.880 2.550 5.410 16.810
CV(%) 9.810 7.260 10.350 11.850 9.880 7.890
Recovery
The recovery of Rat SDF-1α was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 94-107 99
EDTA plasma (n=8) 84-100 91
Cell culture media (n=8) 94-105 100
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Rat SDF-1α and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 91-103 95-112 92-104
Average 98 102 97
1:4 Range 89-106 93-104 101-115
Average 97 99 109
1:8 Range 99-110 96-112 85-95
Average 105 102 90
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Rat SDF-1α. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Rat SDF-1α and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Rat SDF-1α, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Rat SDF-1α. You can calculate the concentration of Rat SDF-1α in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 24772
Uniport ID Q9QZD1
Research Area Cancer , Cardiovascular , Immunology , Microbiology
Rat SDF-1α(Stromal Cell Derived Factor 1α)CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定