Human ADAM9 (A Disintegrin And Metalloprotease 9) CLIA Kit (E-CL-H0221)

Human ADAM9 (A Disintegrin And Metalloprotease 9) CLIA Kit - 1
  • Human ADAM9 (A Disintegrin And Metalloprotease 9) CLIA Kit - 1
  • Human ADAM9 (A Disintegrin And Metalloprotease 9) CLIA Kit - 2
  • Human ADAM9 (A Disintegrin And Metalloprotease 9) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 9.38 pg/mL
Detection Range 15.63-1000 pg/mL
Sample Volume 100 μL
Total Assay Time 4 h 30 min
Reactivity Human
Specificity This kit recognizes Human ADAM9 in samples.No significant cross-reactivity or interference between Human ADAM9 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
1000 54885.000 55725.000 55305.000 55272
500 21796.000 23994.000 22895.000 22862
250 11491.000 10325.000 10908.000 10875
125 4779.000 5221.000 5000.000 4967
62.5 2665.000 2347.000 2506.000 2473
31.25 1346.000 1336.000 1341.000 1308
15.63 680.000 762.000 721.000 688
0 32.000 34.000 33.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 49.850 130.590 455.320 52.140 136.150 441.870
Standard deviation 4.970 14.590 40.660 5.130 11.290 27.440
CV(%) 9.970 11.170 8.930 9.840 8.290 6.210
Recovery
The recovery of Human ADAM9 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 96-109 101
EDTA plasma (n=8) 97-114 105
Cell culture media (n=8) 89-102 94
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human ADAM9 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 87-98 87-101 98-113
Average 93 94 106
1:4 Range 92-104 96-110 90-103
Average 98 101 97
1:8 Range 97-110 99-115 84-100
Average 104 105 91
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human ADAM9. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human ADAM9 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human ADAM9, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human ADAM9. You can calculate the concentration of Human ADAM9 in the samples by comparing the RLU value of the samples to the standard curve.
Human ADAM9 (A Disintegrin And Metalloprotease 9) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}