Human ApoB (Apolipoprotein B) CLIA Kit (E-CL-H0377)

Human ApoB (Apolipoprotein B) CLIA Kit - 1
  • Human ApoB (Apolipoprotein B) CLIA Kit - 1
  • Human ApoB (Apolipoprotein B) CLIA Kit - 2
  • Human ApoB (Apolipoprotein B) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 9.38 ng/mL
Detection Range 15.63-1000 ng/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human ApoB in samples.No significant cross-reactivity or interference between Human ApoB and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
ng/mL OD1 OD2 Mean OD Corrected OD
1000 56043.000 59603.000 57823.000 57798
500 24051.000 24745.000 24398.000 24373
250 11198.000 11072.000 11135.000 11110
125 4538.000 5278.000 4908.000 4883
62.5 2638.000 2618.000 2628.000 2603
31.25 1492.000 1300.000 1396.000 1371
15.63 699.000 735.000 717.000 692
0 24.000 26.000 25.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(ng/mL) 54.880 133.550 375.200 51.940 142.520 342.800
Standard deviation 5.830 12.270 29.340 5.980 14.920 29.760
CV(%) 10.620 9.190 7.820 11.510 10.470 8.680
Recovery
The recovery of Human ApoB was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 88-104 95
EDTA plasma (n=8) 96-108 103
Cell culture media (n=8) 90-103 96
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human ApoB and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 90-107 95-110 96-110
Average 98 101 102
1:4 Range 86-100 101-112 92-107
Average 91 106 98
1:8 Range 88-101 100-114 92-105
Average 93 107 97
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human ApoB. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human ApoB and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human ApoB, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human ApoB. You can calculate the concentration of Human ApoB in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 338
Uniport ID P04114
Research Area Cancer , Cardiovascular , Metabolism , Signal Transduction
Human ApoB (Apolipoprotein B) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.