Facebook
Toll-free:1-888-852-8623

All categories

  • All categories
  • Flow Cytometry Antibodies
  • Immunoassays
  • Antibodies and Reagents
  • Cell Health and Metabolism Assay Kits
  • Cell Isolation Products
  • Cell Identification Kits
  • Proteins and Peptides
  • Cell Culture
Please enter the item number/product keyword!
Keyword cannot be empty !
INSERT SYMBOLS:
  • α
  • β
  • γ
  • δ
  • ε
  • ζ
  • η
  • θ
  • κ
  • μ
  • ω
  • σ
  • τ
  • λ
  • ⅩⅢ
  • ⅩⅢ
  • ⅩⅣ
  • ⅩⅤ
  • ⅩⅦ
  • ⅩⅧ
  • UP ↑

Human bFGF/FGF2 (Basic Fibroblast Growth Factor) CLIA Kit (E-CL-H0394)

Human bFGF/FGF2 (Basic Fibroblast Growth Factor) CLIA Kit - 1
  • Human bFGF/FGF2 (Basic Fibroblast Growth Factor) CLIA Kit - 1
  • Human bFGF/FGF2 (Basic Fibroblast Growth Factor) CLIA Kit - 2
  • Human bFGF/FGF2 (Basic Fibroblast Growth Factor) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 7.5 pg/mL
Detection Range 12.5-800 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human bFGF/FGF2 in samples.No significant cross-reactivity or interference between Human bFGF/FGF2 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
800 52504.000 59152.000 55828.000 55796
400 21449.000 23233.000 22341.000 22309
200 11186.000 9690.000 10438.000 10406
100 4930.000 5046.000 4988.000 4956
50 2310.000 2260.000 2285.000 2253
25 1209.000 1047.000 1128.000 1096
12.5 507.000 573.000 540.000 508
0 31.000 33.000 32.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 39.160 84.030 280.730 39.060 77.900 290.550
Standard deviation 4.970 5.980 32.060 4.660 8.370 20.340
CV(%) 12.690 7.120 11.420 11.930 10.740 7.000
Recovery
The recovery of Human bFGF was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 91-103 97
EDTA plasma (n=8) 98-115 105
Cell culture media (n=8) 99-117 107
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human bFGF and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 90-106 85-100 93-107
Average 97 91 99
1:4 Range 93-106 94-109 97-112
Average 98 101 105
1:8 Range 93-107 99-113 94-108
Average 99 105 100
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human bFGF/FGF2. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human bFGF/FGF2 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human bFGF/FGF2, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human bFGF/FGF2. You can calculate the concentration of Human bFGF/FGF2 in the samples by comparing the RLU value of the samples to the standard curve.
Human bFGF/FGF2 (Basic Fibroblast Growth Factor) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定