Human CNR1 (Cannabinoid Receptor 1, Brain) CLIA Kit (E-CL-H0486)

Human CNR1 (Cannabinoid Receptor 1, Brain) CLIA Kit - 1
  • Human CNR1 (Cannabinoid Receptor 1, Brain) CLIA Kit - 1
  • Human CNR1 (Cannabinoid Receptor 1, Brain) CLIA Kit - 2
  • Human CNR1 (Cannabinoid Receptor 1, Brain) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 18.75 pg/mL
Detection Range 31.25-2000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human CNR1 in samples.No significant cross-reactivity or interference between Human CNR1 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
2000 55927.000 61001.000 58464.000 58430
1000 25240.000 26398.000 25819.000 25785
500 12834.000 10982.000 11908.000 11874
250 4939.000 5887.000 5413.000 5379
125 2911.000 2605.000 2758.000 2724
62.5 1381.000 1345.000 1363.000 1329
31.25 689.000 767.000 728.000 694
0 33.000 35.000 34.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 101.580 241.950 857.530 92.170 236.520 805.010
Standard deviation 10.490 26.400 77.260 8.230 28.220 65.610
CV(%) 10.330 10.910 9.010 8.930 11.930 8.150
Recovery
The recovery of Human CNR1 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 93-107 99
EDTA plasma (n=8) 93-106 99
Cell culture media (n=8) 97-113 105
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human CNR1 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 102-119 88-104 92-106
Average 109 95 98
1:4 Range 94-105 100-119 87-98
Average 100 108 92
1:8 Range 98-115 97-114 94-108
Average 106 104 99
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human CNR1. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human CNR1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human CNR1, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human CNR1. You can calculate the concentration of Human CNR1 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 1268
Uniport ID P21554
Research Area Cancer , Cardiovascular , Metabolism , Neuroscience , Signaltransduction
Human CNR1 (Cannabinoid Receptor 1, Brain) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定