Human CTSK (Cathepsin K) CLIA Kit (E-CL-H0507)

Human CTSK (Cathepsin K) CLIA Kit - 1
  • Human CTSK (Cathepsin K) CLIA Kit - 1
  • Human CTSK (Cathepsin K) CLIA Kit - 2
  • Human CTSK (Cathepsin K) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 18.75 pg/mL
Detection Range 31.25-2000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human CTSK in samples.No significant cross-reactivity or interference between Human CTSK and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
2000 54134.000 65816.000 59975.000 59948
1000 22856.000 25824.000 24340.000 24313
500 11951.000 10189.000 11070.000 11043
250 5169.000 5601.000 5385.000 5358
125 2858.000 2606.000 2732.000 2705
62.5 1412.000 1308.000 1360.000 1333
31.25 704.000 756.000 730.000 703
0 26.000 28.000 27.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 107.500 247.330 853.640 115.290 243.360 802.140
Standard deviation 13.440 18.100 56.600 14.720 23.650 56.550
CV(%) 12.500 7.320 6.630 12.770 9.720 7.050
Recovery
The recovery of Human CTSK was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 87-99 94
EDTA plasma (n=8) 87-98 92
Cell culture media (n=8) 91-107 99
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human CTSK and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 100-117 89-101 101-116
Average 108 94 107
1:4 Range 104-116 100-117 89-105
Average 110 107 97
1:8 Range 94-105 86-100 97-110
Average 99 93 103
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human CTSK. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human CTSK and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human CTSK, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human CTSK. You can calculate the concentration of Human CTSK in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 1513
Uniport ID P43235
Research Area Cancer , Cell Biology , Cardiovascular , Neuroscience , Signal Transduction
Human CTSK (Cathepsin K) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定