Human FGF21 (Fibroblast Growth Factor 21) CLIA Kit (E-CL-H0074)

Human FGF21 (Fibroblast Growth Factor 21) CLIA Kit - 1
  • Human FGF21 (Fibroblast Growth Factor 21) CLIA Kit - 1
  • Human FGF21 (Fibroblast Growth Factor 21) CLIA Kit - 2
  • Human FGF21 (Fibroblast Growth Factor 21) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 7.5 pg/mL
Detection Range 12.5-800 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human FGF21 in samples.No significant cross-reactivity or interference between Human FGF21 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
800 52307.000 58791.000 55549.000 55524
400 22546.000 24006.000 23276.000 23251
200 11502.000 9822.000 10662.000 10637
100 4516.000 4720.000 4618.000 4593
50 2501.000 2139.000 2320.000 2295
25 1130.000 1090.000 1110.000 1085
12.5 516.000 568.000 542.000 517
0 25.000 25.000 25.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 39.320 81.850 297.580 39.700 80.400 316.220
Standard deviation 4.680 7.070 23.480 3.500 7.260 28.620
CV(%) 11.900 8.640 7.890 8.820 9.030 9.050
Recovery
The recovery of Human FGF21 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 85-99 92
EDTA plasma (n=8) 99-113 106
Cell culture media (n=8) 97-113 104
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human FGF21 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 88-105 97-112 95-107
Average 96 103 100
1:4 Range 95-112 96-112 101-116
Average 102 104 108
1:8 Range 83-95 103-115 98-113
Average 90 108 105
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human FGF21. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human FGF21 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human FGF21, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human FGF21. You can calculate the concentration of Human FGF21 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 26291
Uniport ID Q9NSA1
Research Area Cancer , Cardiovascular , Signal Transduction
Human FGF21 (Fibroblast Growth Factor 21) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定