Human GAD1 (Glutamate Decarboxylase 1) CLIA Kit (E-CL-H1400)

Human GAD1 (Glutamate Decarboxylase 1) CLIA Kit - 1
  • Human GAD1 (Glutamate Decarboxylase 1) CLIA Kit - 1
  • Human GAD1 (Glutamate Decarboxylase 1) CLIA Kit - 2
  • Human GAD1 (Glutamate Decarboxylase 1) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 0.19 ng/mL
Detection Range 0.31-20 ng/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human GAD1 in samples.No significant cross-reactivity or interference between Human GAD1 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
ng/mL OD1 OD2 Mean OD Corrected OD
20 34670.000 39038.000 36854.000 36833
10 12866.000 13682.000 13274.000 13253
5 5803.000 5445.000 5624.000 5603
2.5 2373.000 2793.000 2583.000 2562
1.25 1507.000 1319.000 1413.000 1392
0.63 939.000 821.000 880.000 859
0.31 646.000 656.000 651.000 630
0 20.000 22.000 21.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(ng/mL) 0.990 2.400 9.360 1.010 2.230 10.040
Standard deviation 0.120 0.280 0.900 0.100 0.160 1.130
CV(%) 12.120 11.670 9.620 9.900 7.170 11.250
Recovery
The recovery of Human GAD1 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 95-113 103
EDTA plasma (n=8) 98-114 105
Cell culture media (n=8) 102-114 108
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human GAD1 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 101-121 85-101 98-111
Average 110 92 104
1:4 Range 99-111 98-110 97-113
Average 105 104 105
1:8 Range 101-115 95-107 86-97
Average 108 101 91
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human GAD1. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human GAD1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human GAD1, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human GAD1. You can calculate the concentration of Human GAD1 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 2571
Uniport ID Q99259
Research Area Cancer , Metabolism , Immunology , Signal Transduction
Human GAD1 (Glutamate Decarboxylase 1) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定