Human IDE (Insulin Degrading Enzyme) CLIA Kit (E-CL-H0349)

Human IDE (Insulin Degrading Enzyme) CLIA Kit - 1
  • Human IDE (Insulin Degrading Enzyme) CLIA Kit - 1
  • Human IDE (Insulin Degrading Enzyme) CLIA Kit - 2
  • Human IDE (Insulin Degrading Enzyme) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 9.38 pg/mL
Detection Range 15.63-1000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human IDE in samples.No significant cross-reactivity or interference between Human IDE and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
1000 50336.000 59630.000 54983.000 54949
500 23425.000 23677.000 23551.000 23517
250 10885.000 10521.000 10703.000 10669
125 4728.000 5454.000 5091.000 5057
62.5 2590.000 2342.000 2466.000 2432
31.25 1405.000 1209.000 1307.000 1273
15.63 737.000 769.000 753.000 719
0 33.000 35.000 34.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 49.470 111.520 391.000 48.690 108.490 395.330
Standard deviation 6.330 9.320 25.420 4.270 10.660 43.370
CV(%) 12.800 8.360 6.500 8.770 9.830 10.970
Recovery
The recovery of Human IDE was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 86-99 92
EDTA plasma (n=8) 98-113 104
Cell culture media (n=8) 88-101 94
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human IDE and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 86-97 94-108 91-105
Average 91 99 98
1:4 Range 103-117 87-101 101-116
Average 109 93 107
1:8 Range 91-103 98-111 87-101
Average 96 104 92
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human IDE. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human IDE and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IDE, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human IDE. You can calculate the concentration of Human IDE in the samples by comparing the RLU value of the samples to the standard curve.
Human IDE (Insulin Degrading Enzyme) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定