Human IL-1α (Interleukin 1 Alpha) CLIA Kit (E-CL-H0087)

Human IL-1α (Interleukin 1 Alpha) CLIA Kit - 1
  • Human IL-1α (Interleukin 1 Alpha) CLIA Kit - 1
  • Human IL-1α (Interleukin 1 Alpha) CLIA Kit - 2
  • Human IL-1α (Interleukin 1 Alpha) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 2.34 pg/mL
Detection Range 3.91-250 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human IL-1α in samples.No significant cross-reactivity or interference between Human IL-1α and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
250 47548.000 54050.000 50799.000 50770
125 19796.000 20076.000 19936.000 19907
62.5 9766.000 8770.000 9268.000 9239
31.25 4014.000 4140.000 4077.000 4048
15.63 2246.000 1992.000 2119.000 2090
7.81 1129.000 1079.000 1104.000 1075
3.91 604.000 674.000 639.000 610
0 28.000 30.000 29.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 13.260 20.840 98.370 12.950 21.470 101.310
Standard deviation 1.650 2.330 7.220 1.200 2.060 9.840
CV(%) 12.440 11.180 7.340 9.270 9.590 9.710
Recovery
The recovery of Human IL-1α was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 88-100 93
EDTA plasma (n=8) 94-109 100
Cell culture media (n=8) 90-106 97
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human IL-1α and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 98-113 99-110 92-107
Average 104 104 98
1:4 Range 93-106 98-115 94-107
Average 99 105 100
1:8 Range 96-107 88-102 100-113
Average 102 94 108
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human IL-1α. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human IL-1α and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IL-1α, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human IL-1α. You can calculate the concentration of Human IL-1α in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 3552
Uniport ID P01583
Research Area Cancer , Metabolism , Immunology
Human IL-1α (Interleukin 1 Alpha) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定