Human IL-17 (Interleukin 17) CLIA Kit (E-CL-H0104)

Human IL-17 (Interleukin 17) CLIA Kit - 1
  • Human IL-17 (Interleukin 17) CLIA Kit - 1
  • Human IL-17 (Interleukin 17) CLIA Kit - 2
  • Human IL-17 (Interleukin 17) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 7.5 pg/mL
Detection Range 12.5-800 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human IL-17 in samples.No significant cross-reactivity or interference between Human IL-17 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
800 57424.000 58108.000 57766.000 57740
400 20937.000 25527.000 23232.000 23206
200 9990.000 9554.000 9772.000 9746
100 4682.000 5154.000 4918.000 4892
50 2505.000 2245.000 2375.000 2349
25 1138.000 1066.000 1102.000 1076
12.5 532.000 576.000 554.000 528
0 25.000 27.000 26.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 42.680 66.990 285.270 44.380 65.340 282.260
Standard deviation 4.080 7.600 31.890 4.290 6.610 26.760
CV(%) 9.560 11.340 11.180 9.670 10.120 9.480
Recovery
The recovery of Human IL-17 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 97-114 104
EDTA plasma (n=8) 99-114 105
Cell culture media (n=8) 89-102 95
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human IL-17 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 91-107 101-118 88-103
Average 98 108 95
1:4 Range 95-108 91-108 87-103
Average 101 99 94
1:8 Range 86-99 86-96 88-103
Average 91 91 96
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human IL-17. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human IL-17 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IL-17, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human IL-17. You can calculate the concentration of Human IL-17 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 3605
Uniport ID Q16552
Research Area Cancer , Immunology
Human IL-17 (Interleukin 17) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定