Human IL-35 (Interleukin 35) CLIA Kit (E-CL-H1436)

Human IL-35 (Interleukin 35) CLIA Kit - 1
  • Human IL-35 (Interleukin 35) CLIA Kit - 1
  • Human IL-35 (Interleukin 35) CLIA Kit - 2
  • Human IL-35 (Interleukin 35) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 4.69 pg/mL
Detection Range 7.81-500 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human IL-35 in samples.No significant cross-reactivity or interference between Human IL-35 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
500 55183.000 57977.000 56580.000 56561
250 20226.000 23390.000 21808.000 21789
125 10153.000 10077.000 10115.000 10096
62.5 4487.000 4941.000 4714.000 4695
31.25 2431.000 2227.000 2329.000 2310
15.63 1402.000 1222.000 1312.000 1293
7.81 771.000 851.000 811.000 792
0 19.000 19.000 19.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 23.050 59.290 188.360 20.930 61.000 178.660
Standard deviation 1.890 4.400 20.150 2.150 6.810 13.700
CV(%) 8.200 7.420 10.700 10.270 11.160 7.670
Recovery
The recovery of Human IL-35 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 92-107 99
EDTA plasma (n=8) 96-110 102
Cell culture media (n=8) 101-113 107
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human IL-35 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 94-111 102-114 93-104
Average 102 107 98
1:4 Range 99-112 95-108 91-104
Average 106 102 98
1:8 Range 88-101 83-97 87-99
Average 94 90 92
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human IL-35. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human IL-35 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IL-35, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human IL-35. You can calculate the concentration of Human IL-35 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 10148, 3592
Uniport ID P29459 Q14213
Research Area Cancer , Immunology
Human IL-35 (Interleukin 35) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定