Human PAF (Platelet Activating Factor) CLIA Kit (E-CL-H1304)

Human PAF (Platelet Activating Factor) CLIA Kit - 1
  • Human PAF (Platelet Activating Factor) CLIA Kit - 1
  • Human PAF (Platelet Activating Factor) CLIA Kit - 2
  • Human PAF (Platelet Activating Factor) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 37.5 pg/mL
Detection Range 62.5-4000 pg/mL
Sample Volume 100 μL
Total Assay Time 4 h 30 min
Reactivity Human
Specificity This kit recognizes Human PAF in samples.No significant cross-reactivity or interference between Human PAF and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
4000 55185.000 67169.000 61177.000 61155
2000 27175.000 32013.000 29594.000 29572
1000 14056.000 13916.000 13986.000 13964
500 6905.000 7231.000 7068.000 7046
250 3674.000 3302.000 3488.000 3466
125 1682.000 1560.000 1621.000 1599
62.5 756.000 760.000 758.000 736
0 21.000 23.000 22.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 221.410 602.650 1403.590 200.310 564.240 1276.200
Standard deviation 21.810 70.270 160.990 26.020 61.730 118.560
CV(%) 9.850 11.660 11.470 12.990 10.940 9.290
Recovery
The recovery of Human PAF was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 92-106 97
EDTA plasma (n=8) 90-106 96
Cell culture media (n=8) 98-112 106
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human PAF and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 91-107 91-108 87-100
Average 98 99 93
1:4 Range 89-103 89-104 97-114
Average 95 96 105
1:8 Range 103-115 101-115 88-101
Average 109 108 94
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human PAF. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human PAF and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human PAF, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human PAF. You can calculate the concentration of Human PAF in the samples by comparing the RLU value of the samples to the standard curve.
Human PAF (Platelet Activating Factor) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定