Facebook
Toll-free:1-888-852-8623

All categories

  • All categories
  • Flow Cytometry Antibodies
  • Immunoassays
  • Antibodies and Reagents
  • Cell Health and Metabolism Assay Kits
  • Cell Isolation Products
  • Cell Identification Kits
  • Proteins and Peptides
  • Cell Culture
Please enter the item number/product keyword!
Keyword cannot be empty !
INSERT SYMBOLS:
  • α
  • β
  • γ
  • δ
  • ε
  • ζ
  • η
  • θ
  • κ
  • μ
  • ω
  • σ
  • τ
  • λ
  • ⅩⅢ
  • ⅩⅢ
  • ⅩⅣ
  • ⅩⅤ
  • ⅩⅦ
  • ⅩⅧ
  • UP ↑

Human PYGL (Glycogen Phosphorylase, Liver) CLIA Kit (E-CL-H0857)

Human PYGL (Glycogen Phosphorylase, Liver) CLIA Kit - 1
  • Human PYGL (Glycogen Phosphorylase, Liver) CLIA Kit - 1
  • Human PYGL (Glycogen Phosphorylase, Liver) CLIA Kit - 2
  • Human PYGL (Glycogen Phosphorylase, Liver) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 18.75 pg/mL
Detection Range 31.25-2000 pg/mL
Sample Volume 100 μL
Total Assay Time 4 h 30 min
Reactivity Human
Specificity This kit recognizes Human PYGL in samples.No significant cross-reactivity or interference between Human PYGL and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
2000 57831.000 64275.000 61053.000 61026
1000 24364.000 25396.000 24880.000 24853
500 12012.000 11780.000 11896.000 11869
250 5054.000 5686.000 5370.000 5343
125 2864.000 2630.000 2747.000 2720
62.5 1437.000 1403.000 1420.000 1393
31.25 697.000 741.000 719.000 692
0 26.000 28.000 27.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 102.700 256.040 730.140 107.040 261.880 668.250
Standard deviation 9.240 20.640 63.810 12.350 29.930 66.160
CV(%) 9.000 8.060 8.740 11.540 11.430 9.900
Recovery
The recovery of Human PYGL was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 90-103 97
EDTA plasma (n=8) 93-108 99
Cell culture media (n=8) 95-110 102
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human PYGL and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 94-107 102-119 94-105
Average 100 109 99
1:4 Range 102-119 102-118 87-103
Average 108 110 94
1:8 Range 98-110 88-101 85-101
Average 104 93 92
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human PYGL. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human PYGL and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human PYGL, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Human PYGL. You can calculate the concentration of Human PYGL in the samples by comparing the RLU value of the samples to the standard curve.
Human PYGL (Glycogen Phosphorylase, Liver) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定