Mouse COL1 (Collagen Type I) CLIA Kit (E-CL-M0218)

All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 37.5 pg/mL
Detection Range 62.5-4000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Mouse
Specificity This kit recognizes Mouse COL1 in samples.No significant cross-reactivity or interference between Mouse COL1 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
4000 59732.000 65556.000 62644.000 62604
2000 26712.000 31072.000 28892.000 28852
1000 14295.000 13589.000 13942.000 13902
500 6296.000 7210.000 6753.000 6713
250 3658.000 3208.000 3433.000 3393
125 1743.000 1527.000 1635.000 1595
62.5 726.000 788.000 757.000 717
0 39.000 41.000 40.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Mouse COL1 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Mouse COL1 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 210.030 387.120 1685.040 202.640 357.470 1817.140
Standard deviation 17.810 37.320 107.170 24.920 40.470 137.190
CV(%) 8.480 9.640 6.360 12.300 11.320 7.550
Recovery
The recovery of Mouse COL1 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 92-109 100
EDTA plasma (n=8) 101-117 107
Cell culture media (n=8) 95-110 101
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Mouse COL1 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 89-104 91-107 84-98
Average 95 97 91
1:4 Range 87-102 96-112 87-102
Average 93 102 94
1:8 Range 102-117 95-109 97-109
Average 109 101 102
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Mouse COL1. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse COL1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse COL1, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Mouse COL1. You can calculate the concentration of Mouse COL1 in the samples by comparing the RLU value of the samples to the standard curve.
Uniport ID P11087(COL1A1)
Research Area SignalTransduction , Stemcells
Mouse COL1 (Collagen Type I) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定