Mouse VEGF-D (Vascular Endothelial cell Growth Factor D) CLIA Kit (E-CL-M0680)

Mouse VEGF-D (Vascular Endothelial cell Growth Factor D) CLIA Kit - 1
  • Mouse VEGF-D (Vascular Endothelial cell Growth Factor D) CLIA Kit - 1
  • Mouse VEGF-D (Vascular Endothelial cell Growth Factor D) CLIA Kit - 2
  • Mouse VEGF-D (Vascular Endothelial cell Growth Factor D) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 7.5 pg/mL
Detection Range 12.5-800 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Mouse
Specificity This kit recognizes Mouse VEGF-D in samples.No significant cross-reactivity or interference between Mouse VEGF-D and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
800 51756.000 62128.000 56942.000 56909
400 20730.000 24118.000 22424.000 22391
200 10723.000 10211.000 10467.000 10434
100 4725.000 4759.000 4742.000 4709
50 2245.000 2211.000 2228.000 2195
25 1230.000 1048.000 1139.000 1106
12.5 538.000 560.000 549.000 516
0 32.000 34.000 33.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 36.570 70.700 336.350 36.700 67.990 316.300
Standard deviation 3.550 6.820 26.540 3.770 6.540 33.720
CV(%) 9.710 9.650 7.890 10.270 9.620 10.660
Recovery
The recovery of Mouse VEGF-D was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 90-104 95
EDTA plasma (n=8) 99-113 105
Cell culture media (n=8) 93-106 100
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Mouse VEGF-D and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 103-118 102-119 92-104
Average 109 109 99
1:4 Range 88-100 101-117 102-117
Average 95 108 108
1:8 Range 91-105 98-111 92-104
Average 98 103 97
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Mouse VEGF-D. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse VEGF-D and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse VEGF-D, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Mouse VEGF-D. You can calculate the concentration of Mouse VEGF-D in the samples by comparing the RLU value of the samples to the standard curve.
Uniport ID P97946
Research Area Cancer , Cardiovascular , Signal Transduction
Mouse VEGF-D (Vascular Endothelial cell Growth Factor D) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定