Rat DRD3 (Dopamine Receptor D3) CLIA Kit (E-CL-R0221)

Rat DRD3 (Dopamine Receptor D3) CLIA Kit - 1
  • Rat DRD3 (Dopamine Receptor D3) CLIA Kit - 1
  • Rat DRD3 (Dopamine Receptor D3) CLIA Kit - 2
  • Rat DRD3 (Dopamine Receptor D3) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 37.5 pg/mL
Detection Range 62.5-4000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Rat
Specificity This kit recognizes Rat DRD3 in samples.No significant cross-reactivity or interference between Rat DRD3 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
4000 62251.000 63293.000 62772.000 62748
2000 27585.000 33333.000 30459.000 30435
1000 14886.000 13026.000 13956.000 13932
500 6735.000 6739.000 6737.000 6713
250 3536.000 3492.000 3514.000 3490
125 1637.000 1535.000 1586.000 1562
62.5 727.000 799.000 763.000 739
0 24.000 24.000 24.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 219.850 510.660 1716.980 216.220 557.030 1623.010
Standard deviation 23.280 56.270 174.790 24.710 58.880 148.990
CV(%) 10.590 11.020 10.180 11.430 10.570 9.180
Recovery
The recovery of Rat DRD3 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 97-112 103
EDTA plasma (n=8) 88-101 95
Cell culture media (n=8) 99-115 107
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Rat DRD3 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 99-115 86-100 99-113
Average 105 92 107
1:4 Range 88-104 96-110 89-102
Average 96 102 96
1:8 Range 88-100 90-105 90-102
Average 93 96 95
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Rat DRD3. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Rat DRD3 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Rat DRD3, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Rat DRD3. You can calculate the concentration of Rat DRD3 in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 29238
Uniport ID P19020
Research Area Cancer , Neuroscience
Rat DRD3 (Dopamine Receptor D3) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定