Facebook
Toll-free:1-888-852-8623

All categories

  • All categories
  • Flow Cytometry Antibodies
  • Immunoassays
  • Antibodies and Reagents
  • Cell Health and Metabolism Assay Kits
  • Cell Isolation Products
  • Cell Identification Kits
  • Proteins and Peptides
  • Cell Culture
Please enter the item number/product keyword!
Keyword cannot be empty !
INSERT SYMBOLS:
  • α
  • β
  • γ
  • δ
  • ε
  • ζ
  • η
  • θ
  • κ
  • μ
  • ω
  • σ
  • τ
  • λ
  • ⅩⅢ
  • ⅩⅢ
  • ⅩⅣ
  • ⅩⅤ
  • ⅩⅦ
  • ⅩⅧ
  • UP ↑

Rat OC/BGP (Osteocalcin) CLIA Kit (E-CL-R0169)

Rat OC/BGP (Osteocalcin) CLIA Kit - 1
  • Rat OC/BGP (Osteocalcin) CLIA Kit - 1
  • Rat OC/BGP (Osteocalcin) CLIA Kit - 2
  • Rat OC/BGP (Osteocalcin) CLIA Kit - 3
All Size Price Qty
96T $ 682.00
- +
48T $ 546.00
- +
24T $ 180.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 0.09 ng/mL
Detection Range 0.16-10 ng/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Rat
Specificity This kit recognizes Rat OC/BGP in samples.No significant cross-reactivity or interference between Rat OC/BGP and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Chemiluminescence
Assay Type Sandwich-CLIA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃/-20℃
Expiration Date 6 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
ng/mL OD1 OD2 Mean OD Corrected OD
10 31581.000 34813.000 33197.000 33158
5 16881.000 17047.000 16964.000 16925
2.5 8435.000 7113.000 7774.000 7735
1.25 3517.000 3805.000 3661.000 3622
0.63 2037.000 1767.000 1902.000 1863
0.31 836.000 832.000 834.000 795
0.16 364.000 406.000 385.000 346
0 38.000 40.000 39.000 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(ng/mL) 0.480 0.910 4.910 0.440 0.980 4.530
Standard deviation 0.060 0.100 0.490 0.040 0.110 0.280
CV(%) 12.500 10.990 9.980 9.090 11.220 6.180
Recovery
The recovery of Rat OC was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 86-97 91
EDTA plasma (n=8) 95-110 101
Cell culture media (n=8) 93-110 100
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Rat OC and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 87-101 91-106 94-108
Average 93 97 99
1:4 Range 89-102 100-115 91-109
Average 95 106 99
1:8 Range 87-99 87-101 88-102
Average 92 92 93
This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Rat OC/BGP. Samples or Standards are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Rat OC/BGP and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Rat OC/BGP, biotinylated detection antibody and Avidin-HRP conjugate will appear luminescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer at 425nm. The RLU value is positively associated with the concentration of Rat OC/BGP. You can calculate the concentration of Rat OC/BGP in the samples by comparing the RLU value of the samples to the standard curve.
Gene ID 25295
Uniport ID P04640
Research Area Cancer , Signal Transduction , Stem Cells
Rat OC/BGP (Osteocalcin) CLIA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定