Human F1+2(Prothrombin Fragment 1+2) ELISA Kit (E-EL-H1793)

Human F1+2(Prothrombin Fragment 1+2) ELISA Kit - 1
  • Human F1+2(Prothrombin Fragment 1+2) ELISA Kit - 1
  • Human F1+2(Prothrombin Fragment 1+2) ELISA Kit - 2
  • Human F1+2(Prothrombin Fragment 1+2) ELISA Kit - 3
All Size Price Qty
96T $ 609.00
- +
48T $ 487.00
- +
24T $ 150.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 28.13 pg/mL
Detection Range 46.88-3000 pg/mL
Sample Volume 100 μL
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes Human F1+2 in samples. No significant cross-reactivity or interference between Human F1+2 and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Colorimetric method, ELISA, Sandwich
Assay Type Sandwich-ELISA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃
Expiration Date 12 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
3000 2.245 2.141 2.193 2.112
1500 1.616 1.566 1.591 1.510
750 0.911 0.869 0.890 0.809
375 0.437 0.455 0.446 0.365
187.5 0.303 0.293 0.298 0.217
93.75 0.176 0.178 0.177 0.096
46.88 0.133 0.133 0.133 0.052
0 0.082 0.080 0.081 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 145.620 420.430 1189.830 144.180 425.590 1199.670
Standard deviation 9.540 24.130 79.960 8.750 31.070 86.020
CV(%) 6.550 5.740 6.720 6.070 7.300 7.170
Recovery
The recovery of Human F1+2 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Cell culture media (n=8) 88-103 95
Serum (n=8) 88-99 94
EDTA plasma (n=8) 90-102 97
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human F1+2 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 88-100 88-98 98-111
Average 94 93 105
1:4 Range 87-98 82-93 92-104
Average 93 88 98
1:8 Range 84-95 84-94 87-102
Average 90 89 94
Stability
Each kit batch is subjected to accelerated stability testing and real-time stability monitoring. Sample performance is evaluated after storage at 37 °C for 10 days to assess the impact of elevated temperature on assay reliability and reagent integrity.
/ Variation range of 37°C mean
concentration / 2-8°C mean
Sample 1(n=16) 97.18-102.64
Sample 2(n=16) 92.26-98.74
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human F1+2. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human F1+2 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human F1+2, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human F1+2. You can calculate the concentration of Human F1+2 in the samples by comparing the OD of the samples to the standard curve.
Gene ID 2147
Uniport ID P00734
Research Area Cardiovascular
Human F1+2(Prothrombin Fragment 1+2) ELISA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

Species

{{relatedElisaPage.numTotal}} Results

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定