Facebook
Toll-free:1-888-852-8623

All categories

  • All categories
  • Flow Cytometry Antibodies
  • Immunoassays
  • Antibodies and Reagents
  • Cell Health and Metabolism Assay Kits
  • Cell Isolation Products
  • Cell Identification Kits
  • Proteins and Peptides
  • Cell Culture
Please enter the item number/product keyword!
Keyword cannot be empty !
INSERT SYMBOLS:
  • α
  • β
  • γ
  • δ
  • ε
  • ζ
  • η
  • θ
  • κ
  • μ
  • ω
  • σ
  • τ
  • λ
  • ⅩⅢ
  • ⅩⅢ
  • ⅩⅣ
  • ⅩⅤ
  • ⅩⅦ
  • ⅩⅧ
  • UP ↑
Micro-Sample

Mini Sample Mouse RETN (Resistin) ELISA Kit (E-MSEL-M0010)

Mini Sample Mouse RETN (Resistin) ELISA Kit - 1
  • Mini Sample Mouse RETN (Resistin) ELISA Kit - 1
  • Mini Sample Mouse RETN (Resistin) ELISA Kit - 2
  • Mini Sample Mouse RETN (Resistin) ELISA Kit - 3
All Size Price Qty
96T $ 625.00
- +
48T $ 500.00
- +
24T $ 150.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary

The sample size of MS series ELISA kit is only 1/4 of that of the traditional sandwich ELISA kit, which can effectively solve the problem of insufficient sample size.

Sensitivity 16.44 pg/mL
Detection Range 31.25-2000 pg/mL
Sample Volume 25 μL
Total Assay Time 3 h 30 min
Reactivity Mouse
Specificity This kit recognizes Mouse RETN in samples.No significant cross-reactivity or interference between Mouse RETN and analogues was observed
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Colorimetric method, ELISA, Sandwich
Assay Type Sandwich-ELISA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃
Expiration Date 12 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
2000 2.369 2.333 2.351 2.261
1000 1.811 1.777 1.794 1.704
500 1.102 1.078 1.090 1
250 0.516 0.532 0.524 0.434
125 0.312 0.316 0.314 0.224
62.5 0.196 0.202 0.199 0.109
31.25 0.145 0.143 0.144 0.054
0 0.091 0.089 0.090 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 104.150 244.980 760.450 100.960 254.310 763.350
Standard deviation 6.350 12.490 40.760 7.660 18.640 46.030
CV(%) 6.100 5.100 5.360 7.590 7.330 6.030
Recovery
The recovery of Mouse RETN was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 86-99 91
EDTA plasma (n=8) 91-102 97
Cell culture media (n=8) 90-106 97
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Mouse RETN and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 96-114 95-108 93-110
Average 104 101 101
1:4 Range 92-108 95-113 97-109
Average 100 103 102
1:8 Range 96-114 97-110 93-109
Average 104 104 100
Stability
Each kit batch is subjected to accelerated stability testing and real-time stability monitoring. Sample performance is evaluated after storage at 37 °C for 10 days to assess the impact of elevated temperature on assay reliability and reagent integrity.
/ Variation range of 37°C mean
concentration / 2-8°C mean
Sample 1(n=16) 92.24-94.30
Sample 2(n=16) 91.41-108.99
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse RETN. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse RETN and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse RETN, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Mouse RETN. You can calculate the concentration of Mouse RETN in the samples by comparing the OD of the samples to the standard curve.
Gene ID 57264
Uniport ID Q99P87
Research Area Cardiovascular , Metabolism , Signaltransduction
Mini Sample Mouse RETN (Resistin) ELISA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

Species

{{relatedElisaPage.numTotal}} Results

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定