Mini Sample Human MCP-1 ( Monocyte Chemotactic Protein 1 ) ELISA Kit (E-MSEL-H0013)

Mini Sample Human MCP-1 ( Monocyte Chemotactic Protein 1 ) ELISA Kit - 1
  • Mini Sample Human MCP-1 ( Monocyte Chemotactic Protein 1 ) ELISA Kit - 1
  • Mini Sample Human MCP-1 ( Monocyte Chemotactic Protein 1 ) ELISA Kit - 2
  • Mini Sample Human MCP-1 ( Monocyte Chemotactic Protein 1 ) ELISA Kit - 3
All Size Price Qty
96T $ 510.00
- +
48T $ 408.00
- +
24T $ 150.00
- +
96T*5 Inquire /
96T*10 Inquire /
Add to cart

For research use only.

Product Summary
Sensitivity 9.38 pg/mL
Detection Range 15.63-1000 pg/mL
Sample Volume 25 μL
Manual Operation Time 1 h
Total Assay Time 3 h 30 min
Reactivity Human
Specificity This kit recognizes MCP-1 in samples.No significant cross-reactivity or interference between MCP-1 and analogues was observed.
Recovery 80%-120%
Sample Type Serum, plasma and other biological fluids
Detection Method Colorimetric method, ELISA, Sandwich
Assay Type Sandwich-ELISA
Size 96T / 48T / 24T / 96T*5 / 96T*10
Storage 2-8℃
Expiration Date 12 months
Typical data
The following data was generated by the Quality Control Department, under controlled laboratory conditions (ambient temperature: 18-25 °C, relative humidity: 35-75%) using standardized procedures (TMB reaction at 37 °C in the dark for 15 minutes, followed by termination and OD measurement). These values are provided for reference only. Actual results may vary due to differences in laboratory conditions, operator technique, and equipment. Users are required to generate a standard curve using their own experimental data.
pg/mL OD1 OD2 Mean OD Corrected OD
1000 2.247 2.191 2.219 2.125
500 1.414 1.372 1.393 1.299
250 0.846 0.832 0.839 0.745
125 0.450 0.478 0.464 0.370
62.5 0.288 0.274 0.281 0.187
31.25 0.182 0.188 0.185 0.091
15.63 0.143 0.145 0.144 0.050
0 0.093 0.095 0.094 -
Precision
Intra-assay Precision (Within-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested 20 times on a single plate.
Inter-assay Precision (Between-run Precision): Three samples representing low, mid, and high concentrations of Human IL-6 were tested on three separate plates, with 20 replicates per plate, to assess variability among assays.
/ Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
Numbers 20 20 20 20 20 20
Mean(pg/mL) 45.510 154.440 491.510 42.490 148.400 452.800
Standard deviation 1.840 10.040 32.830 3.210 11.120 32.510
CV(%) 4.040 6.500 6.680 7.550 7.490 7.180
Recovery
The recovery of Human MCP-1 was evaluated by spiking samples at low, mid, and high concentrations across the assay range in various sample matrices.
The assay performance was assessed by comparing the measured concentrations to the expected spiked amounts to determine the percent recovery.
Sample Type Range (%) Average Recovery(%)
Serum (n=8) 92-104 97
EDTA plasma (n=8) 84-98 91
Cell culture media (n=8) 91-102 97
Linearity
Linearity of the assay was evaluated by spiking samples with high concentrations of Human MCP-1 and performing serial dilutions using Standard & Sample Diluent to produce concentrations spanning the assay's dynamic range. The measured values were then compared to the expected concentrations to assess the linearity of response.
/ / Cell culture media
(n=5)
EDTA plasma
(n=5)
Serum
(n=5)
1:2 Range 95-112 98-112 95-112
Average 102 105 103
1:4 Range 92-107 99-113 92-105
Average 97 105 98
1:8 Range 89-106 100-114 95-110
Average 97 106 101
Stability
Each kit batch is subjected to accelerated stability testing and real-time stability monitoring. Sample performance is evaluated after storage at 37 °C for 10 days to assess the impact of elevated temperature on assay reliability and reagent integrity.
/ Variation range of 37°C mean
concentration / 2-8°C mean
Sample 1(n=16) 88.21-100.52
Sample 2(n=16) 91.11-91.87
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to MCP-1 . Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for MCP-1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain MCP-1 , biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of MCP-1 . You can calculate the concentration of MCP-1 in the samples by comparing the OD of the samples to the standard curve.
MCP-1, also known as CCL2, is a member of the CC chemokine family. It is widely expressed in endothelial cells, smooth muscle cells and monocytes in response to several atherogenic stimulants such as CD40 ligand, platelet derived growth factor (PDGF), interleukin-1β (IL-1β) and oxidized low density lipoprotein.
Gene Alias CCL2
Gene ID 6347
Uniport ID P13500
Protein Alias CCL2
Research Area Cell Biology , Neuroscience
Mini Sample Human MCP-1 ( Monocyte Chemotactic Protein 1 ) ELISA Kit - procedures
  • IF:{{item.impact}}

    Journal:{{item.journal}} ({{item.year}})

    DOI:{{item.doi}}

    Reactivity:{{item.species}}

    Sample Type:{{item.sample_type}}

Species

{{relatedElisaPage.numTotal}} Results

  • Q{{(FAQpage.currentPage - 1)*pageSize+index+1}}:{{item.name}}

No files available. Please enter the correct identification code in the search area to query the relevant documents.

empty
Product Information
Title Status Valid until Action
{{v.文件名}} {{v.status_str}} {{v.timestamp_str}} Download /

*Note:Please download the relevant files within the validity period. After expiration, file preview and download will not be available on the page.

确定