Total Protein (TP) Colorimetric Assay Kit (With standard, BCA method)

    • Biochemical Assay Kit-Elabscience
    <
    >
    • Biochemical Assay Kit-Elabscience
    • Biochemical Assay Kit-Elabscience

      Catalog number:E-BC-K318

      Size:
      • 48T
      • 96T
      Qty:
      - +
      Price: $120

      Detection method: Colorimetric method

      Detection instrument: Microplate reader

      Valid period: 6 months

      Lead Time: 7~10 daysWelcome to order from local distributors.

      Add to cart Compare Bulk request Manual


      Application

      This kit can measure protein content of a variety of animal serum, plasma, culture cells, tissue and other samples.

       

      Detection significance

      The BCA protein concentration Kit is an ideal protein quantification method which is superior to the Lowry method. This method is fast and sensitive, stable and reliable to different types of protein with small variation coefficient, which is greatly favored by professionals. The BCA method is not affected by the chemicals for most samples.

       

      Detection principle

      Cu2+ can be reduced to Cu+ by protein in alkaline condition. Cu+ can combine with BCA reagent and form purple complex, which has a maximum absorption peak at 562 nm. The absorbance value is proportional to the protein concentration. Therefore, the protein concentration can be calculated according to the OD value.


      Experimental instruments

      Test tube, Micropipettor, Vortex mixer, Water bath, Microplate reader (562 nm), Centrifuge



      Sample pretreatment

      1. Serum sample:

      Fresh blood was collected and placed at 25 for 30 min to clot the blood. Centrifuge the sample at 4 for 15 min at 2000 ×g, the upper yellowish clear liquid was taken as serum. Place the serum on ice for detection. If not detected on the same day, stored the serum at-80, which can be stored for a month.

       

      2. Plasma sample:

      The fresh blood was added into the test tube containing anticoagulant and mixed upside down. Centrifuge the sample at 4 for 10 min at 700~1000 ×g, the upper yellowish transparent liquid was taken as the plasma, and the middle white interference layer (white blood cells and platelets) could not be absorbed. Place the plasma on ice for detection. If not detected on the same day, stored the serum at-80, which can be stored for a month.

       

      3. 10% tissue homogenate sample:

      Accurately weigh the tissue sample, add 9 times the volume of PBS (0.01 M, pH7~7.4) according to the ratio of Weight (g): Volume (mL) =1:9. Mechanical homogenate the sample in ice water bath. Centrifuge at 10000 g for 10 min, then take the supernatant and preserve it on ice for detection. Meanwhile, determine the protein concentration of supernatant (E-BC-K318, E-BC-K168, E-BC-K165).

       

      4. Culture cell sample:

      Wash the cells with PBS (0.01 M, pH7~7.4) for 1~2 times. Centrifuge at 1000 g for 10 min and then discard the supernatant and keep the cell sediment. Add PBS at a ratio of cell number (106): PBS (μL) =1: 300-500. Sonicate or grind with hand-operated in ice water bath. Centrifuge at 10000 g for 10 min, then take the supernatant and preserve it on ice for detection. Meanwhile, determine the protein concentration of supernatant (E-BC-K318, E-BC-K168, E-BC-K165).

       

      Note: The detection ranges is from 16.5 to 1000μg/ml, please take 1-2 preliminary experiments before sample tests, and then diluted the samples with normal saline.

      Recommended dilution multiple:

      Serum (plasma) sample: 50-70 times;

      Animal tissue homogenate: 8-12 times;

      Plant tissue homogenate: 3-6 times;

      Cell sample: 2-3 times.

       

      Operation steps

      1. Blank tube: add 20 μL of normal saline or PBS.

      Standard tube: add 20 μL of 563 μg/mL Standard.

      Sample tube: add 20 μL of tested Samples.

      2.  Add 200 μL of BCA working solution to the wells of Step 1.

      3. Oscillate for 20 s to mix fully and incubate at 37℃ for 30 min.

      4. Measure the OD value of each well at 562 nm with Microplate Reader.

       

      Note: The following operating table could be as a reference.


       

      Blank well

      Standard well

      Sample well

      Saline or PBS (μL)

      20

       

       

      563 μg/mL Standard (μL)

       

      20

       

      Samples (μL)

       

       

      20

      BCA working solution (μL)

      200

      200

      200

      Oscillate for 20 s to mix fully and incubate at 37 for 30 min. Measure the OD values of each well at 562 nm with Microplate Reader.


      Technical parameter

      1. The sensitivity of the kit is 16.5 μg/mL.

      2. The intra CV is 2.2% and the inter CV is 4.5%.

      3. The recovery of the kit is 100%.

      4. The detection range of the kit is 16.5-1000 μg/mL.

       

      Notes

      1. This kit is for research use only.

      2. Instructions should be followed strictly, changes of operation may result in unreliable results.

      3. The validity of kit is 6 months.

      4.  Do not use components from different batches of kit.

      5. Control the incubation time strictly.

      6. The concentration of the sample protein must be diluted to 1000 μg / mL or less with normal saline, and it will show a good linear range below this concentration.

      7. Prevent the formulation of bubbles when adding the reagents to the microplate.